| SYMPTOM |
POSSIBLE
CAUSE |
REMEDY |
A—ENGINE
FAILS TO FIRE WHEN COLD
|
1. Inadequate
or no fuel supply
|
1. Slacken filter plug
on A.E.D. Crank engine, fuel should leak from
to A.E.D. plug.
If no fuel, check system leading to A.E.D. If fuel is present,
tighten plug. Then carry out check 2 under symptom A.
|
| 2. Inadequate
or no fuel supply from A.E.D. to inlet manifold |
2. Crank engine for several
seconds. Remove pipe at inlet manifold.
If fuel is present, it indicates that the A.E.D. is satisfactory
and the cause for failure to start must be traced to some other
source. If no fuel is present this indicates a faulty A.E.D.
Proceed as follows:
(i) Check that main valve is open by inserting probe down centre hole—press
down fully, probe should return approximately .030 in. (0,7S mm) when
released.
NOTE: This check is not applicable when temperature is above 350C.
(ii) Remove float chamber lid, check that needle valve and float are
free to move.
If there is no fuel in float chamber check that fuel filter is clean
and that ample fuel is being delivered to the unit. See Operation
N/Q 21A. |
| B—ENGINE
FIRES BUT FAILS TO KEEP RUNNING WHEN COLD |
1. Sticking
or faulty needle valve or float
|
1. Check the following:
i) Remove vent pipe from float chamber and check that fuel
is not discharged when cranking engine.
(ii) If fuel is discharged check operation of needle valve and float
as detailed under ‘A’ item 2 (ii). |
| 2. Faulty
air flap valve |
2. Remove air inlet elbow
on A.E.D. and check that flap valve is
free to move, and return under spring load. If jammed dismantle
unit and rectify. |
| 3. Inadequate
fuel supply |
3. Check for lack of fuel
as detailed under ‘A’ Item land ‘A’ Item 2
(ii). |
| 4. Faulty
jet needle diaphragm |
4. Remove cap for
needle diaphragm, then withdraw diaphragm and
check for damage. Replace faulty parts. Reassemble as detailed
under Operation NIQ 20A.
If diaphragm is satisfactory
a complete overhaul of the unit is
indicated. |
| C—ENGINE
FAILS TO START WHEN HOT OR PART WARM OR FIRES AND FAILS TO KEEP
RUNNING
|
1. Incorrect
starting procedure |
1. Crank engine and open
throttle slightly, if unit is badly over-
choked open throttle fully do not over-rev. If engine starts but
fails to keep running carry out check 4’ below. |
| 2. Leaks
from pipe layout |
2. Ensure’ that
all pipes are correctly fitted, particularly the pipe
between the hot air pick-up and the A.E.D. Rectify as necessary
ensuring air tight joints.
NOTE: Air leaks at the hot air entry will result in excessive fuel
consumption and the possibility of stalling at certain
engine temperatures. |
| 3. Before
carrying Out further checks ensure that the A.E.D.is not the cause
of failure tostart |
3. Remove pipe from A.E,D.
to inlet manifold. Blank off aperture
in manifold. If engine starts this indicates A.E.D. is at fault, and
further checks must be made. However, if engine is flooded it may
be necessary to crank for several seconds, with slightly open
throttle before the engine will fire. |
| 4. Sticking
or faulty needle valve or float |
4. Check the following:
(I) Remove vent pipe from float chamber and check that fuel is
not discharged when cranking engine.
(ii) If fuel is discharged check operation of needle valve and float.
Remove float chamber lid, check that needle valve and float are free
to move.
if there is no fuel in float chamber check that fuel filter is clean
and that ample fuel is being delivered to the unit. See Operation
NIQ 2lA. |
| 5. Main valve
faulty. This fault and the checking procedure applies only when engine
really hot. |
5. Check that main valve
is completely closed by inserting probe
down centre hole, press down fully, probe should not return which
is indicates that valve is fully seated. If probe returns under spring
pressure dismantle unit and rectity. |
| 6. Incorrect,
needle movement |
6. Check movement of jet
needle by inserting probe down hollow
grub screw, push down fully, if engine is at its normal working
temperature probe should not return, if partly warm, probe
should return approximately .015 in. (0,40 mm). |
| 7. Faulty
needle diaphragm |
7. Carry out checks as
detailed under ‘B’ Item 4. |